Fluke 754 Overview

Youre going to love what weve done with your calibrator.

The 754 is everything you needed and wished for in the 744 and more! HART communication capability is combined to deliver an integrated communicating calibrator. This rugged, reliable tool is ideal for calibrating, maintaining, and troubleshooting HART and other instrumentation.

The 754 does the work of several tools sourcing, simulating and measuring pressure, temperature, and electrical signals in one rugged, hand-held device. Youll fall in love with the new improved graphical screen, Li-Ion battery for longer life, USB port and new accessories to complete the package.

For documentation, the 754 automates calibration procedures and captures your data. And, of course, it helps you meet rigorous standards like ISO 9000, FDA, EPA and OSHA regulations.

HART Capabilities

The 754 is designed to take on nearly all the day-to-day tasks you now perform with a separate communicator. In fact, it offers most of the communication capabilities of a HART communicator.

Each Fluke 754 Includes

  • Three sets of stackable test leads
  • Three sets of TP220 test probes with three sets of extended tooth alligator clips
  • Two sets AC280 hook clips
  • BP7240 Li-ion battery pack
  • BC7240 battery charger
  • C799 field soft case
  • USB communication cable
  • Getting started guide
  • Instruction manual on CDROM
  • NIST traceable certificate of calibration
  • DPCTrack™ 2 sample software
  • HART communication cable

>

Measurement Accuracy
Voltage DC Range/Resolution 1 Year 2 Years
100.000 mV 0.02% 0.005 mV 0.03% 0.005 mV
3.00000 V 0.02% 0.00005 V 0.03% 0.00005 V
30.0000 V 0.02% 0.0005 V 0.03% 0.0005 V
300.00 V 0.05% 0.05 V 0.07% 0.05 V
Voltage AC 3.000 V (40 Hz to 500 Hz) / 0.001 V 0.5% 0.002 V 1.0% 0.004 V
30.00 V (40 Hz to 500 Hz) / 0.01 V 0.5% 0.02 V 1.0% 0.04 V
300.0 V (40 Hz to 500 Hz) / 0.1 V 0.5% 0.2 V 1.0% 0.2 V
Current DC 30.000 mA 0.01% 5 uA 0.015% 7 uA
110.00 mA 0.01% 20 uA 0.015% 30 uA
Resistance 10.000 Ω 0.05% 50 mΩ 0.07% 70 mΩ
100.00 Ω 0.05% 50 mΩ 0.07% 70 mΩ
1.0000 kΩ 0.05% 500 mΩ 0.07% 0.5 Ω
10.000 kΩ 0.1% 10 Ω 0.15% 15 Ω
Frequency 1.00 to 110.00 Hz / 0.01 Hz 0.05 Hz
110.1 to 1100.0 Hz / 0.1 Hz 0.5 Hz
1.101 to 11.000 kHz / 0.001 kHz 0.005 kHz
11.01 to 50.00 kHz / 0.01 kHz 0.05 kHz
Source Accuracy
1 Year 2 Years
Voltage DC 100.000 mV 0.01% 0.005 mV 0.015% 0.005 mV
1.00000 V 0.01% 0.00005 V 0.015% 0.0005 V
15.0000 V 0.01% 0.0005 V 0.015% 0.0005 V
Current DC 22.000 mA (source) 0.01% 0.003 mA 0.02% 0.003 mA
Current sink (simulate) 0.02% 0.007 mA 0.04% 0.007 mA
Resistance 10.000 Ω 0.01% 10 mΩ 0.015% 15 mΩ
100.00 Ω 0.01% 20 mΩ 0.015% 30 mΩ
1.0000 kΩ 0.02% 0.2 Ω 0.03% 0.3 Ω
10.000 kΩ 0.02% 3 Ω 0.03% 5 Ω
Frequency 0.1 to 10.99 Hz 0.01 Hz
0.01 to 10.99 Hz 0.01 Hz
11.00 to 109.99 Hz 0.1 Hz
110.0 to 1099.9 Hz 0.1 Hz
1.100 to 21.999 kHz 0.002 kHz
22.000 to 50.000 kHz 0.005 kHz
Technical Data
Data log functions Measure functions Voltage, current, resistance, frequency, temperature, pressure
Reading rate 1, 2, 5, 10, 20, 30, or 60 readings/minute
Maximum record length 8000 readings (7980 for 30 or 60 readings/minute)
Ramp functions Source functions Voltage, current, resistance, frequency, temperature
Rate 4 steps/second
Trip detect Continuity or voltage (continuity detection not available when sourcing current)
Loop power function Voltage Selectable, 26 V
Accuracy 10%, 18 V minimum at 22 mA
Maximum current 25 mA, short circuit protected
Maximum input voltage 50 V DC
Step functions Source functions Voltage, current, resistance, frequency, temperature
Manual step Selectable step, change with arrow buttons
Autostep Fully programmable for function, start delay, stepvalue, time per step, repeat
Environmental Specifications
Operating temperature -10°C to 50°C
Storage temperature -20°C to 60°C
Dust/water resistance Meets IP52, IEC 529
Operating altitude 3000 m above mean sea level (9842 ft)
Safety Specifications
Agency approvals CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993
Mechanical and General Specifications
Size 136 x 245 x 63 mm (5.4 x 9.6 x 2.5 in)
Weight 1.2 kg (2.7 lb)
Batteries Internal Battery Pack Li-ion: 7.2V,4400mAh, 30 Wh
Battery life &gt,8 hours typical
Battery replacement Replace without opening calibrator, no tools required
Side port connections Pressure module connector
USB Connector to interface to your PC
Digital instrument (HART) connector
Connection for optional battery charger/eliminator
Data storage capacity 1 week of calibration procedures results
90 day specifications The standard specification interval for the 750 Seriesare 1 and 2 years.
Typical 90 day measurement and source accuracy can be estimated by dividing the one year % of reading or % of output specifications by 2.
Floor specifications, expressed as % of full scale or counts or ohms remain constant.
Temperature, Resistance Temperature Detectors
Degrees or % of reading Type (α) Range °C Measure °C1
1 year 2 year
100 Ω Pt (385) -200 to 100 0.07°C 0.14°C
100 to 800 0.02% 0.05°C 0.04% 0.10°C
200 Ω Pt (385) -200 to 100 0.07°C 0.14°C
100 to 630 0.02% 0.05°C 0.04% 0.10°C
500 Ω Pt (385) -200 to 100 0.07°C 0.14°C
100 to 630 0.02% 0.05°C 0.04% 0.10°C
1000 Ω Pt (385) -200 to 100 0.07°C 0.14°C
100 to 630 0.02% 0.05°C 0.04% 0.10°C
100 Ω Pt (3916) -200 to 100 0.07°C 0.14°C
100 to 630 0.02% 0.05°C 0.04% 0.10°C
100 Ω Pt (3926) -200 to 100 0.08°C 0.16°C
100 to 630 0.02% 0.06°C 0.04% 0.12°C
10 Ω Cu (427) -100 to 260 0.2°C 0.4°C
120 Ω Ni (672) -80 to 260 0.1°C 0.2°C
Source current Source °C Allowable current2
1 year 2 year
1 mA 0.05°C 0.10°C 0.1 mA to 10 mA
0.0125% 0.04°C 0.025% 0.08°C
500 μA 0.06°C 0.12°C 0.1 mA to 1 mA
0.017% 0.05°C 0.034% 0.10°C
250 μA 0.06°C 0.12°C 0.1 mA to 1 mA
0.017% 0.05°C 0.034% 0.10°C
150 μA 0.06 C 0.12 C 0.1 mA to 1 mA
0.017% 0.05°C 0.034% 0.10°C
1 mA 0.05°C 0.10°C 0.1 mA to 10 mA
0.0125% 0.04°C 0.025% 0.08°C
1 mA 0.05°C 0.10°C 0.1 mA to 10 mA
0.0125% 0.04°C 0.025% 0.08°C
3 mA 0.2°C 0.4°C 0.1 mA to 10 mA
1 mA 0.04°C 0.08°C 0.1 mA to 10 mA
1. For two and three-wire RTD measurements, add 0.4°C to the specifications.
2. Supports pulsed transmitters and PLCs with pulse times as short as 1 ms
Temperature, Thermocouples
Type Source °C Measure °C Source °C
1 year 2 years 1 year 2 years
E -250 to -200 1.3 2 0.6 0.9
-200 to -100 0.5 0.8 0.3 0.4
-100 to 600 0.3 0.4 0.3 0.4
600 to 1000 0.4 0.6 0.2 0.3
N -200 to -100 1.0 1.5 0.6 0.9
-100 to 900 0.5 0.8 0.5 0.8
900 to 1300 0.6 0.9 0.3 0.4
J -210 to -100 0.6 0.9 0.3 0.4
-100 to 800 0.3 0.4 0.2 0.3
800 to 1200 0.5 0.8 0.3 0.3
K -200 to -100 0.7 1 0.4 0.6
-100 to 400 0.3 0.4 0.3 0.4
400 to 1200 0.5 0.8 0.3 0.4
1200 to 1372 0.7 1 0.3 0.4
T -250 to -200 1.7 2.5 0.9 1.4
-200 to 0 0.6 0.9 0.4 0.6
0 to 400 0.3 0.4 0.3 0.4
B 600 to 800 1.3 2 1 1.5
800 to 1000 1 1.5 0.8 1.2
1000 to 1820 0.9 1.3 0.8 1.2
R -20 to 0 2.3 2.8 1.2 1.8
0 to 100 1.5 2.2 1.1 1.7
100 to 1767 1 1.5 0.9 1.4
S -20 to 0 2.3 2.8 1.2 1.8
0 to 200 1.5 2.1 1.1 1.7
200 to 1400 0.9 1.4 0.9 1.4
1400 to 1767 1.1 1.7 1 1.5
C 0 to 800 0.6 0.9 0.6 0.9
800 to 1200 0.8 1.2 0.7 1
1200 to 1800 1.1 1.6 0.9 1.4
1800 to 2316 2 3 1.3 2
L -200 to -100 0.6 0.9 0.3 0.4
-100 to 800 0.3 0.4 0.2 0.3
800 to 900 0.5 0.8 0.2 0.3
U -200 to 0 0.6 0.9 0.4 0.6
0 to 600 0.3 0.4 0.3 0.4
BP 0 to 1000 1 1.5 0.4 0.6
1000 to 2000 1.6 2.4 0.6 0.9
2000 to 2500 2 3 0.8 1.2
XK -200 to 300 0.2 0.3 0.2 0.5
300 to 800 0.4 0.6 0.3 0.6

Additional information

Brand

Fluke

Be the first to review “Fluke 754: The Ultimate Multifunction Calibrator for HART and Instrumentation Calibration”

Your email address will not be published. Required fields are marked *

Reviews

There are no reviews yet.

Main Menu

Fluke 754: The Ultimate Multifunction Calibrator for HART and Instrumentation Calibration

$4,421.00

Add to Cart